Simultaneous detection of salivary cortisol and cortisone using an automated high-throughput sample preparation method for LC-MS/MS

Open AccessPublished:February 20, 2022DOI:https://doi.org/10.1016/j.slast.2022.01.006

      Abstract

      High-Performance Liquid Chromatography-Tandem Mass Spectrometry has emerged triumphant over the years as a reliable and high throughput clinical instrument to assess different metabolic irregularities. One of the latest applications of LC-MS/MS has been in the field of quantification of glucocorticoids, such as cortisol and cortisone from saliva, that can be an indicator of abnormalities such as Congenital Adrenal Hyperplasia (CAH), Cushing's Syndrome, and Addison's disease. We have developed and validated an LC-MS/MS-based assay for simultaneous detection of cortisol and cortisone in human saliva, which requires only 20 µL of sample, to measure cortisol across a 0.5 – 70 ng/mL range, and cortisone across a 1.2– 100 ng/mL range, respectively. The developed method exhibits linearity of R2>0.99, for both analytes, inclusive of both MRMs, and a percent coefficient of variation that is less than or equal to 20%. Using dilute-and-shoot for sample preparation, we have exhibited sample accuracy of 100±20 for the assay calibrators, and integrated Needle Wash Solvent Chemistry for minimal sample carryover, making it adaptable for crucial for potential diagnostic use. We have exhibited a method that is simplistic, specific, and highly automatable on liquid handling platforms, such as the JANUS® G3 Workstation. With our innovation, we have introduced a potential to test 81-unknown samples in singlicate within an on-deck plating time of fewer than four minutes. We performed additional verification studies, including an accelerated stability study and a freeze-thaw study showcasing the potential long-term usability of our proposed prototype kit. Overall, this work presents an optimized LC-MS/MS method with automated sample preparation that is ready to be utilized for cortisol-cortisone detection for clinical diagnostic contexts related to Cushing's Disease, among other adrenal and endocrine disorders. 

      1. Introduction

      Glucocorticoids (GC), an important class of lipophilic, fat-soluble steroid hormones, regulate a wide variety of physiological functions. GCs modulate numerous processes including mood and cognitive function, growth, reproduction, and development, immune responses, as well as cardiovascular regulation [
      • Timmermans S.
      • Souffriau J.
      • Libert C.
      A general introduction to glucocorticoid biology.
      ,
      • Stephens M.A.C.
      • Wand G.
      Stress and the HPA axi: Role of glucocorticoids in alcohol dependence.
      ,
      • Ehlert U.
      • Gaab J.
      • Heinrichs M.
      Psychoneuroendocrinological contributions to the etiology of depression, posttraumatic stress disorder and stress-related bodily disorders: the role of the hypothalamus–pituitary–adrenal axis.
      ]. As a result of this functional diversity, human GC nuclear receptors are expressed ubiquitously throughout the body and in all organs and tissues [
      • Nicolaides N.C.
      • Galata Z.
      • Kino T.
      • Chrousos G.P.
      • Charmandari E.
      The human glucocorticoid receptor: molecular basis of biologic function.
      ]. Two notable examples of endogenous glucocorticoids are cortisol (CF) and its inactive form, cortisone (CE).
      Cortisol, widely known as the stress hormone, is synthesized in, and released from the adrenal cortex, resulting in elevated heart and respiratory rates. These are characteristic of the stress responses, which are triggered because of stressful stimuli that cause the hypothalamus activation of the hypothalamus-pituitary-adrenal (HPA) axis [
      • Ramamoorthy S.
      • Cidlowski J.A.
      Corticosteroids: mechanisms of action in health and disease.
      ]. Elevated levels of cortisol trigger metabolic pathways in the liver, pancreas, and adipose tissue that are responsible for maintaining glucose and protein balance throughout the body [
      • Hannibal K.E.
      • Bishop M.D.
      Chronic stress, cortisol dysfunction and pain: a psychoneuroendocrine rationale for stress management in pain rehabilitation.
      ]. High cortisol signals for increased gluconeogenesis and decreased glycogen synthesis and in the liver and for lipolysis in adipose tissue. Moreover, in the presence of high levels of cortisol, muscle cells decrease glucose uptake and consumption and shift towards increased protein degradation to provide the glucogenic amino acids necessary for gluconeogenesis [
      • Kuo T.
      • McQueen A.
      • Chen T.-C.
      • Wang J.-C.
      Regulation of glucose homeostasis by glucocorticoids.
      ,
      • Exton J.
      Regulation of gluconeogenesis by glucocorticoids.
      ].
      Dysregulated levels of cortisol and cortisone may arise from abnormalities in glucocorticoid-activating enzymes such as 11 β-hydroxylase and 21-hydroxylase deficiencies. The latter examples can take place in Congenital Adrenal Hyperplasia (CAH), as well as other inflammatory, psychiatric, and metabolic disorders [
      • Olgemöller B.
      • Roscher A.A.
      • Liebl B.
      • Fingerhut R.
      Screening for congenital adrenal hyperplasia: adjustment of 17-hydroxyprogesterone cut-off values to both age and birth weight markedly improves the predictive value.
      ,
      • New M.I.
      Diagnosis and management of congenital adrenal hyperplasia.
      ]. Similarly, Cushing's disease and Addison's disease are both prominent adrenal disorders where clinical testing showcases clear signs of irregular levels of corticosteroids [
      • Yamamoto T.
      Latent adrenal insufficiency: concept, clues to detection and diagnosis.
      ,
      • Mitchell I.C.
      • et al.
      Subclinical Cushing’s syndrome” is not subclinical: improvement after adrenalectomy in 9 patients.
      ]. Circulating cortisol-cortisone can therefore function as biomarkers that allow for the detection and monitoring of chronic stresses, adrenal disorders, and more.
      Herein, we have developed and validated a clinical method for metabolic irregularity assessment that utilizes a simple dilute-and-shoot protocol for the analysis of circulating cortisol-cortisone via Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) in saliva. LC-MS/MS is a clinical analytical technique that has matured over the past decade, becoming the gold standard in many biochemical-diagnostics [
      • Seger C.
      • Salzmann L.
      After another decade: LC–MS/MS became routine in clinical diagnostics.
      ,
      • Erickson D.
      • Singh R.J.
      • Sathananthan A.
      • Vella A.
      • Bryant S.C.
      Late-night salivary cortisol for diagnosis of Cushing’s syndrome by liquid chromatography/tandem mass spectrometry assay.
      ]. LC-MS/MS brings many advantages over other analyte detection alternatives, such as ELISA and immunostaining [
      • Casals G.
      • Hanzu F.A.
      Cortisol measurements in Cushing’s syndrome: immunoassay or mass spectrometry?.
      ]. Among other advantages, LC-MS/MS provides the ability to accurately quantify analytes at low concentrations while maintaining high analyte specificity and exhibits superior sensitivity over other analytical techniques. This is critical in the context of cortisol-cortisone, where high cross-reactivity is frequently observed in immunoassays [
      • Chabre O.
      The difficulties of pseudo-Cushing’s syndrome (or “non-neoplastic hypercortisolism”).
      ,
      • Raff H.
      • Phillips J.M.
      Bedtime salivary cortisol and cortisone by LC-MS/MS in healthy adult subjects: evaluation of sampling time.
      ]. The proposed protocol was optimized to measure salivary analyte levels, as salivary cortisol-cortisone levels are similarly indicative of free cortisol in plasma, serum, and urine [
      • Erickson D.
      • Singh R.J.
      • Sathananthan A.
      • Vella A.
      • Bryant S.C.
      Late-night salivary cortisol for diagnosis of Cushing’s syndrome by liquid chromatography/tandem mass spectrometry assay.
      ]; however, saliva is easier to obtain and carries fewer disadvantages [
      • Do Yup Lee E.K.
      • Choi M.H.
      Technical and clinical aspects of cortisol as a biochemical marker of chronic stress.
      ,
      • Jung C.
      • Greco S.
      • Nguyen H.H.
      • Ho J.T.
      • Lewis J.G.
      • Torpy D.J.
      • Inder W.J.
      Plasma, salivary and urinary cortisol levels following physiological and stress doses of hydrocortisone in normal volunteers.
      ,
      • Papanicolaou Dimitris A.
      • et al.
      Nighttime salivary cortisol: a useful test for the diagnosis of Cushing's syndrome.
      ]. For example, attempts at quantifying free-cortisol from urine samples encounter heightened matrix effects, due to high amounts of endogenous impurities, leading to ion suppression and loss of sensitivity [
      • Jung C.
      • Greco S.
      • Nguyen H.H.
      • Ho J.T.
      • Lewis J.G.
      • Torpy D.J.
      • Inder W.J.
      Plasma, salivary and urinary cortisol levels following physiological and stress doses of hydrocortisone in normal volunteers.
      ,
      • Tudela E.
      • Muñoz G.
      • Muñoz-Guerra J.A.
      Matrix effect marker for multianalyte analysis by LC-MS/MS in biological samples.
      ]. Plasma samples require more complex sample preparation protocols, often necessitating a solid phase extraction step that reduces high-throughput capacity and increases processing costs per sample. Lastly, simultaneously measuring cortisol-cortisone from plasma may potentially yield misleading results [
      • Jung C.
      • Greco S.
      • Nguyen H.H.
      • Ho J.T.
      • Lewis J.G.
      • Torpy D.J.
      • Inder W.J.
      Plasma, salivary and urinary cortisol levels following physiological and stress doses of hydrocortisone in normal volunteers.
      ], due to the 8:1 ratio of cortisol: cortisone in plasma, as opposed to the 1:2 ratio in saliva [
      • Hannibal K.E.
      • Bishop M.D.
      Chronic stress, cortisol dysfunction and pain: a psychoneuroendocrine rationale for stress management in pain rehabilitation.
      ,
      • Bakusic J.
      • De Nys S.
      • Creta M.
      • Godderis L.
      • Duca R.C.
      Study of temporal variability of salivary cortisol and cortisone by LC-MS/MS using a new atmospheric pressure ionization source.
      ]. The 11β-hydroxysteroid dehydrogenase in the salivary glands aids the conversion of unbound cortisol to cortisone, leading to this vast difference in abundance. Therefore, saliva is highly indicated as a diagnostic matrix for rapid diagnosis of diseases affecting the adrenal glands via assessment of circulating glucocorticoids.
      Plasma as a means for detecting free cortisol demands a sample preparation process of a higher degree of complexity due to an often-required solid phase extraction procedure that reduces the high-throughput capacity of the proposed method, as well as increases the cost of processing per sample. Lastly, measuring cortisol from plasma may potentially yield misleading results due to its measurement of the total amount of cortisol, including bound cortisol [
      • Jung C.
      • Greco S.
      • Nguyen H.H.
      • Ho J.T.
      • Lewis J.G.
      • Torpy D.J.
      • Inder W.J.
      Plasma, salivary and urinary cortisol levels following physiological and stress doses of hydrocortisone in normal volunteers.
      ]. Ultimately, this encourages the utilization of saliva as a means for rapid recognition, diagnosis, and treatment of diseases affecting the adrenal glands by accurately evaluating circulating glucocorticoid levels.
      The present work focuses on optimizing streamlined sample preparation and LC-MS/MS analysis method that avoids the need for error-bound extraction steps (liquid-liquid extraction, solid-phase extraction, etc.) often found in other protocols. This reduces the complexity of the sample preparation process, allowing for a more streamlined and high-throughput automation using conventional robotic liquid handlers, with minimal needed human interaction. Overall, as shown in Table 1, the methodology of this work focuses on a rationalized and simplified approach, while maintaining the linearity of analyte detection with relevant clinical ranges of disease. The proposed method exhibits an excellent range of linearity for established salivary cortisol for CAH, dexamethasone suppression, ACTH stimulation, Syndrome of Apparent Mineralocorticoid Excess (SAME), and other metabolic disorders [

      Niewoehner, C. B. (2004). 5. In Endocrine pathophysiology (Second, pp. 85–105). essay, Hayes Barton Press.

      ,
      • Stewart P.M.
      • Corrie J.E.T.
      • Shackleton C.H.L.
      • Edwards C.R.W.
      Syndrome of apparent mineralocorticoid excess. A defect in the cortisol-cortisone shuttle.
      ]. The proposed protocol has the potential to be utilized as a high-throughput research-only kit, that can measure the free cortisol-cortisone in human saliva with great precision and accuracy using only a 20 µL patient sample volume, across the range of 0.5 to 70 ng/mL (CF) and 1.2 to 100 ng/mL (CE) distributed across six calibrators and low and high-Quality Controls (QCs) .
      Table 1a) A comparative study of our method against existing literature. It is important to note that our proposed method targets a more clinically relevant range, unlike the other studies.
      a)
      StepFariha et al. (2021)Bakusic et al. (2019) [
      • Bakusic J.
      • De Nys S.
      • Creta M.
      • Godderis L.
      • Duca R.C.
      Study of temporal variability of salivary cortisol and cortisone by LC-MS/MS using a new atmospheric pressure ionization source.
      ]
      Magda et al. (2017) [
      • Magda B.
      • Dobi Z.
      • Mészáros K.
      • Szabó É.
      • Márta Z.
      • Imre T.
      • et al.
      Charged derivatization and on-line solid phase extraction to measure extremely low cortisol and cortisone levels in human saliva with liquid chromatography–tandem mass spectrometry.
      ]
      Antonelli et al. (2015) [
      • Antonelli G.
      • Ceccato F.
      • Artusi C.
      • Marinova M.
      • Plebani M.
      Salivary cortisol and cortisone by LC–MS/MS: validation, reference intervals and diagnostic accuracy in Cushing’s syndrome.
      ]
      Fustinoni et al. (2013) [
      • Fustinoni S.
      • Polledri E.
      • Mercadante R.
      High-throughput determination of cortisol, cortisone and melatonin in oral fluid by on-line turbulent flow liquid chromatography interfaced with liquid chromatography/tandem mass spectrometry.
      ]
      1Sample Pre-treatment: Thaw samples to room temperature; Centrifuge at 4600 rpm for 10 minutes to collect clear solutionSample Pre-treatment: Centrifuge at 2000 xg for 5 minutes; store at -80 °C; thaw before useSample Pre-treatment: Centrifuge at 2000 xg for 10 minutes; store at -20 °C; thaw before useSample Pre-treatment: Centrifuge at 2000 xg for 10 minutes; store at -80 °C; Centrifuge at 2000 xg for 10minutes; thaw before useSample Pre-treatment: Centrifuge at 3000 rpm for 5 minutes; store at -20 °C (in the dark); thaw before use
      2Sample Conditioning: Sample

      (V=20 µL) + IS (V= 120 µL)

      [Total V= 140 µL]
      Sample Conditioning:

      Sample

      (V=200 µL) + IS (20 µL) + Water (V= 800 µL)

      [Total V= 1020 µL]
      Sample Conditioning: Sample

      (V=250 µL) + IS in ethyl acetate (V= 750 µL)

      [Total V= 1000 µL]
      Sample Conditioning:

      Sample

      (V=300 µL) + IS (V= 30 µL) + Methanol (V= 1000 µL) + Water (V= 1000 µL)

      [Total V= 2330 µL]
      Sample Conditioning:

      Sample

      (V= 500 µL) + IS (V= 5 µL)

      [Total V= 505 µL]
      3Dilute and Shoot: Shake at 800 rpm at 50°C for 10 minutesSPE via PRiME HLB cartridge [V= 1000 µL]LLE: Vortex for 2 minutesSPE via Oasis HLb cartridge

      [V= 250 µL]
      LC-MS/MS (with TurboFlow)
      4LC-MS/MSSample Wash:

      500 µL water:methanol (95:5 v/v)
      5 minutes incubation (room temp.)Sample Wash: 500 µL water:methanol (80:20 v/v)N/A
      5N/AExtraction w/ 500 µL methanol700 µL separationExtraction w/ 250 µL methanolN/A
      6N/AExtraction w/ 500 µL methanolEvaporationLC-MS/MS (with on-line SPE)N/A
      7N/AEvaporationReconstitution: 700 µL ethanol

      [+ Derivatization]
      N/AN/A
      8N/AReconstitution: water:methanol (50:50 v/v) in 0.1% FA (V= 1000 µL)LC-MS/MSN/AN/A
      9N/ALC-MS/MSN/AN/AN/A
      ResultsLoD: CF = 0.5 ng/mL CE = 1.2 ng/mLLoD: CF = 0.02 ng/mL CE = 0.02 ng/mLLoD:

      CF = 0.002 ng/mL

      CE = 0.005 ng/mL
      LoD:

      CF = 0.07 ng/mLCE = 0.04 ng/mL
      LoD:

      CF = 0.10 ng/mL

      CE= 0.10 ng/mL
      b)
      Clinical Ranges of Cortisol-Cortisone in Different Metabolic Disorders
      DisorderCF (Cortisol)CE (Cortisone)Reference(s)
      Congenital Adrenal Hyperplasia (CAH)0.5-2410 nmol/L0.8-134nmol/LBacila, Irina, et al. [
      • Bacila Irina
      • et al.
      Measurement of salivary adrenal-specific androgens as biomarkers of therapy control in 21-hydroxylase deficiency.
      ]
      Dexamethasone Suppression

      (Oral Dexamethasone at 0.015 mg/kg body weight for normal-weight individuals)
      Men:

      24.6 ± 7.4 nmol/L

      Women:

      25.1 ± 23.2 nmol/L
      Not InvestigatedPasquali, Renato, et al. [
      • Pasquali Renato
      • et al.
      Cortisol and ACTH response to oral dexamethasone in obesity and effects of sex, body fat distribution and dexamethasone concentrations: a dose-response study.
      ]
      ACTH Stimulation

      (250μg administration)
      6.7–23.0 nmol/LNot InvestigatedOspina, Naykky Singh, et al. [
      • Ospina Naykky Singh
      • et al.
      ACTH stimulation tests for the diagnosis of adrenal insufficiency: systematic review and meta-analysis.
      ]
      Table 2(Top: Cortisol- Quantifier; Bottom: Cortisone- Quantifier) % CV was used as a measure for precision, determined using 3 replicates per level each day, over a 6-day period (n=18 per level). Samples were found to be within 12% or less for intraday and interday measurements, inclusive of both MRMs for both analytes. The average % accuracy for all calibrators was found to be 100±20, inclusive of both MRMs for both analytes. These values fall well within our ≤20% cut-off value.

      2. Experimental

      2.1 Chemicals and reagents

      Baker Analyzed HPLC grade water, acetonitrile, and methanol (>99.9% purity) were obtained from VWR for the preparation of calibrators, test samples, and the equilibration of the chromatographic column to avoid particulate matter that might potentially affect LC-MS/MS performance. Ampules of primary analytes of study – Cortisol (C21H30O5, MW= 362.46 amu), Cortisone (C21H28O5, MW= 360.44 amu), Cortisol-D4 (internal standard) (C21H26D4O5, MW=366.48 amu) and Cortisone-13C3 (internal standard) (C1813C3H28O5, MW= 363.42 amu) were purchased from Cerilliant Corporation (Round Rock, TX, USA). Reagent grade formic acid was obtained from Sigma-Aldrich. Utak SMX Oral Fluid (Valencia, CA) was used as the base matrix to create the calibrators and QCs.

      2.2 Preparation of calibrators, quality controls, and stock solutions

      The cortisol and cortisone stock solutions were initially used to prepare spike solutions, using a two-step dilution method. This was done to ensure a 2.5% or less volume percentage of methanol across all calibrators and QCs, yielding five spike solutions for cortisol and six for cortisone, respectively. The Utak SMX Oral Fluid was then spiked using these spike solutions to create the calibrators and QCs, covering a concentration range of 0.5-70.0 ng/mL (CF) and 1.2-100.0 ng/mL (CE). Calibrators, as well as QCs, were aliquoted and stored at -80°C before starting any system suitability study. Additionally, a temperature-controlled centrifuge step (4°C, at 4600 RPM) was added to the protocol and investigated as a pre-treatment condition to emphasize solid particulate pellet formation, to prevent system clogging after injection of calibrators and QCs in LC-MS/MS.
      Daily Working Solution (DWS) was prepared in a 50:50 (v/v) solution of HPLC water to HPLC methanol, with the final concentration at 1.7 ng/mL and 1.3 ng/mL for cortisol and cortisone, respectively. The DWS was stored at -20°C and was warmed to room temperature prior to each assay.

      2.3 LC-MS/MS conditions and optimized parameters

      The cortisol and cortisone analytes were diluted 10,000- and 1000-fold respectively, in HPLC-grade methanol before being directly infused into the MS to quantify the parent mass, quantifier fragment, and qualifier fragment masses. This was also repeated with the corresponding analyte internal standards, diluted 100 folds in HPLC-grade methanol. The compound transition data, together with the Entrance Voltage (EV), Collision Cell Energy (CC), and Collision Cell Lens 2 (CCL2) values obtained from Multiple Reaction Monitoring (MRM) optimization, are available in Supplemental Table S1.
      After MRM optimization, the nebulizer gas pressure, source temperature, HSID (hot surface-induced desolvation) temperature, and drying gas pressure were also optimized to obtain the best relative intensity for both Q3 masses of the analytes and their internal standards. After the Q1 and Q3 masses were detected, the EV, CC, and CCL2 values were re-optimized for the best signal of the analyte quantifier, qualifier, and internal standards. The LC-MS/MS system used was a PerkinElmer QSight® 200-series Triple Quadrupole Mass Spectrometer, equipped with an Electron Spray Injection (ESI) source in using positive ion mode. Integrated to the MS was a QSight LX50 Solvent Delivery Module, QSight LX50 Precision Sampling Module, and QSight® LX-50 Column Stability Module for the column temperature control. The column used was a PerkinElmer Brownlee® SPP C-18 column (2.7 µm, C18, 90 Å, 4.6 × 75mm), set to be 40°C for the entire assay run, with mobile phase A (inorganic) and mobile phase B (organic) according to an optimized flow profile to the flow ontable 4. The phase was designed with a linear hold, ranging from 100% A to 95% B, and back to 100% A (phase-flow design is available in Supplemental table S2). For improved carryover mitigation, the Precision Sampling Module needle was put through weak-strong-weak solvent wash cycles, where the weak solvent was 70:30= water: acetonitrile solution and the strong solvent was 100% HPLC acetonitrile. The volume was kept at 250 µL for both wash solvents during the wash cycles. Sample delivery was carried out through a 250 µL syringe, connected to a 10 µL needle and a 20 µL loop, with triple quadruple MS operation at a nebulizer pressure of 200 psi, electrospray voltage of 5850 volts, and a source temperature of 425°C. This produced runs of a total of 2.5 minutes for each sample for simultaneous analyte detection. Afterward, all subsequent chromatographic analyses were performed on the Simplicity™ 3Q (version 1.9) software designed specifically for the QSight instrument.
      We performed a phase test with 100 ng/mL concentration of cortisol and cortisone, dissolved in ratios of organic to inorganic solvents ranging from organic:inorganic=95:5to5:95. In addition, we performed a 4-parameter test (data not reported) to test the solubility of the analytes in organic solvents (acetonitrile and methanol) and their absolute intensities with respect to two different organic solvents used for mobile phase B. Acetonitrile as mobile phase B yielded the strongest relative signal for each calibrator level and methanol was determined to be a better solvent for dissolving the analytes. The formic acid level was determined to be 0.1% for both mobile phases and the flow rate was held steady at 0.6 mL/min throughout the experiment.

      2.4 Sample preparation

      Calibrators, QCs, and samples were thawed at room temperature, followed by centrifuging at 4600 RPM for 10 minutes at 4°C. Avoiding the particulate pellet at the bottom, 20μL of the clear solution from the top was added to each well on a large volume 96-well plate, followed by the addition of 120 μL of the DWS to each well. HPLC water was used as blanks. The plate was properly sealed and shaken using the TriNEST™ microplate shaker at 800 RPM for ten minutes at a controlled temperature of 50°C, before inserting the 96-well plate into the LC-MS/MS autosampler with a Rapid EPS plate seal (BioChromato, Japan) for analysis.

      2.5 Accuracy, precision, and linearity

      In order to assess the validity of the proposed LC-MS/MS method, testing of several parameters indicating the assay sensitivity was performed. This was done in a 6-day study, where each calibrator level and QCs were analyzed in triplicates. To confirm method linearity over the specified target range, calibration curves were evaluated from the analysis of the six different calibrator levels on Simplicity 3Q (Version 1.9) software. A standard curve was generated using linear regression with 1x weighting, where the analyte-to-IS ratios were compared against known calibrator concentrations. The series inter-assay linearity was evaluated from this linear regression, with the reported R-Square value as the primary assessment of assay linearity.
      Similarly, assay precision was measured as the percent coefficient of variation (%CV), which was obtained from the aforementioned calibration curves for each level within run, between run, and total. Sample accuracy was obtained from the automatically reported analysis from Simplicity 3Q software and averaged for all levels across all runs. For each plate, quality control samples (low= 3 ng/mL for cortisol, 6 ng/mL for cortisone; high= 30 ng/mL for cortisol, 50 ng/mL for cortisone) were run together with the calibrators for further validation of the inter-assay precision.

      2.6 Freeze-thaw

      Information on the developed cortisol-cortisone quantification kit's stability was obtained by performing freeze-thaw testing over multiple cycles. Freeze-thaw stability testing was performed on a full prototype kit, comprising the six cortisol-cortisone calibrators, the two quality controls, as well as the internal standard. Cortisol-cortisone kits were produced and subjected to six freeze-thaw cycles via transfer from one of two cold reservoirs (-20°C and -80°C) to room temperature and measuring concentration changes of cortisol-cortisone over the six freeze-thaw cycles by analyzing each calibrator level in triplicate.

      2.7 Assessing sample carry-over

      To confirm a minimal sample carryover for the proposed sample preparation and LC-MS/MS methodology, the highest calibrator level was injected into the system, followed by triplicate injections of the lowest calibrator level. Any deviation in the measurement of the cortisol-cortisone quantities in the lowest level was observed and recorded, with sample carryover accounting for any significant changes in the analyte concentrations due to the first injection of the highest calibrator concentration.

      2.8 Evaluating stability of prepared standards

      To obtain stability information on the proposed cortisol-cortisone diagnostic kit, we performed an accelerated 14-day accelerated stability study in compliance with the Clinical and Laboratory Standards Institute (CLSI) guidelines for stability testing of in vitro diagnostic reagents. Kit calibrators were subjected to extreme temperature changes by being moved across different storage temperatures. Two sets of calibrators were tested for stability, with one being transferred from -80°C to -20°C, and another from -20°C to +30°C; calibrators of the same level were transferred in duplicates. Transfers were staggered and done every two days, for a total of seven transfers over the total period of 14 days.
      After the last transfer on the 14th day, samples were harvested and tested against fresh samples to confirm the threshold of stability. Following this accelerated stability study and the confirmation of analyte retention after sample stressing, the following equation (derived from Arrhenius’ rate equation) was utilized to obtain an estimate for calibrator stability without any degradation. This derivation utilizes the ten-degree rule, where the rate of reaction is said to double for every ten degrees Celsius change in temperature .
      StableDays=(Daysinhightemperature)×(2ΔTemperature10)


      Fig. 1:
      Fig. 1Representative chromatograms showing unspiked salivary matrix (top panel), lowest level of calibrator, L1 (middle panel) and internal standard (bottom panel) for cortisol (left) and cortisone (right). The LoD should be 3:1= signal: baseline for acceptability, and with visible peaks for both analytes when compared to blank, our LoD meets this criterion. The result yielded by the blank matrix was almost identical to HPLC water.

      Equation 1: Modified Arrhenius’ rate equation for measurement of sample stability.

      2.9 Automation of sample preparation protocol

      To further adapt the cortisol-cortisone quantification kit to clinical applications, the developed sample preparation method was automated and tested for improved high-throughput capacity using the JANUS® G3 Workstation (PerkinElmer Corp., MA, U.S.). The sample preparation process previously described was adapted onto the JANUS® workstation using JANUS® Application Assistant and WinPREP® for JANUS®, via the deck arrangement illustrated in Supplemental Fig. 1. To determine the efficacy of this automated sample preparation method, manually prepared samples were compared against automated ones with regards to precision and accuracy.
      The JANUS® G3 Workstation is used to automate the sample preparation. The viscosity of the samples was found comparable to water for all standard levels of cortisol-cortisone in oral fluid. Using this information, performance files for liquids with viscosity near or at that of water were used; the details of the performance files for the Blowout Mode, and the Waste Mode can be found in Supplemental Table S3(a) and S3(b), respectively1). These performance files ensure that the aspirations and dispenses are accurate and precise each time. Depending on our application and plate layouts, we utilized both Blowout and Waste Mode performance files. For a singlicate dispense, such as the calibrators, we used the Blowout Mode, whereas, for dispenses per aspirate ≥ 2, such as an unknown sample to be tested, we used the Waste Mode. The design of this automation protocol was to maximize the number of unknown samples that could be quantified accurately per assay.
      The DWS containing methanol required the addition of a pipette-tip prewetting step and the introduction of transport air gaps into the tip to combat volatile liquid loss. By aspirating and dispensing volume into and out of the pipette tip, the tip becomes ‘prewet’; this action saturates the interior of the pipetting tip with vapors from the solution. When the final aspiration is made, the interaction between the tip and the volatile liquid will no longer create vapor build-up, which is a significant cause of dripping. In addition, air gaps assisted in minimizing the leakage. Aspirating 10 µl of air after initial aspiration and each subsequent dispense was beneficial in combating the dripping of volatile DWS from the pipetting tips.

      3. Results and discussion

      3.1 Blank matrix study

      For every assay run, L0 (unspiked oral fluid) was injected in triplicate into LC-MS/MS system with the goal of ensuring minimal signal noise. The calculated concentration for both cortisol and cortisone were < 0±0.008 ng/mL, or no peak was observed, demonstrating the absence of either analyte in the blank matrix. The criteria of acceptance for LoD is a signal-to-baseline ratio of 3-to-1, which our LoD met for both cortisol-cortisone (Fig. 2).
      Fig. 2:
      Fig. 2Quantifier (top) and Qualifier (bottom) concentrations measured against the theoretical concentrations for cortisol-cortisone. The measured concentrations were used in the generation of the standard curves for all further tests.

      3.2 Phase test and 4-parameter test

      The main purpose of performing the phase test was to obtain the lowest percentage of organic solvent at the end of the final step of the sample preparation, to yield a symmetrical chromatogram with no peak distortion and good signal intensity. An ideal peak should yield a symmetrical Gaussian profile (Equation 2), with minimal peak broadening, exhibiting the highest column efficiency.
      y=1σ2πexp[12(xx¯σ)2];


      where σ=standarddeviation, x¯=meanandx=x¯attheaxisofsymmetry

      Equation 2: Gaussian distribution for chromatogram with no peak distortion.

      This meant that any percentage of organic and inorganic that produced chromatograms with peak fronting and peak tailing were not considered. Even though the water:acetonitrile=55:45 ratio gave the best peak shape, upon inspecting the individual MRM intensities, we observed that water:acetonitrile=35:65 yielded the highest relative intensity for both the analytes . That said, assessing the signal-to-noise ratio for each analyte was also crucial to the method development process. Both analytes (for quantifier and qualifier) for eachinorganic:organic ratio were tested in triplicates and we averaged the MRM intensity and signal-to-noise ratio for each solvent ratio to conclude that water:acetonitrile=60:40 resulted in a Gaussian peak with the best MRM intensity, along with a high signal, and low noise (selected simplified chromatograms are available in Supplemental Fig. S2).
      Fig. 3
      Fig. 3Predicted and experimental concentrations for cortisol and cortisone in the 14-day accelerated stability study. While the predicted trend shows a much sharper decline, experimentally we found both analytes to be stable for a longer time than predicted.

      3.3 True concentration test (QC Study)

      Because the calibrators were prepared using commercially available cortisol-cortisone in methanol, there was a possibility of the concentration of the actual stock solution not matching the stated theoretical value. Since cortisol and cortisone do not have NIST (National Institute of Standards and Technology) certified material, we had to establish a true concentration of calibrators and quality controls for each manufacturing batch, assuming that the stock solution concentrations measured exactly at 1 mg/mL. As each set of calibrators and quality controls were prepared, each level was analyzed in sextuplets and averaged to obtain the ‘true concentration’, to avoid error propagation in precision measurements for the following experiments. Fig. 6 shows both quantifier and qualifier mass data for both analytes and the ‘true concentrations’ against the theoretical concentrations.

      3.5 Linearity, precision, and accuracy

      Cortisol and cortisone calibrators both reported intra and inter-assay precision (%CV) measured at ≤20%. Moreover, for further validation of the results, inter-assay %CV of the QCs of both cortisol-cortisone both reported measurements of ≤10%. Generally, each lab sets the precision threshold for their LC-MS/MS studies, with ±15-20% being standard. Given the absence of NIST-certified materials for cortisol and cortisone, it is challenging to standardize any clinical assay. Hence, we utilized our True Concentration Test to establish the QCs and calibrators such that they can be used to establish the baseline and quantify any QCs for future calibrator manufacturing. The accuracy for the calibrators was within the 100±20% range for both analytes (Table 2). This establishes our method, together with the use of the UTak salivary matrix, to be implemented in clinical laboratory settings for the generation of standard curves for the said glucocorticoid measurements in patient samples. The R-square values were taken as a measure of the linearity, generated in Simplicity 3Q software using linear regression. Linearity was calculated by plotting the mean measured analyte concentrations, as a function of the corresponding known enrichment levels and performing linear regression analysis. For both cortisol and cortisone, the R-square value was greater than 0.99 for both MRMs across all 6 days (Supplemental Table S4). In addition, the precision, linearity, and accuracy studies were performed for cortisol-cortisone quantification kits in two storage temperatures: -80°C (reported) and -20°C (not reported). The results obtained for both temperatures were very similar, with the %CV ≤20% for both cortisol and cortisone (inclusive of both MRMs) at both storage temperatures, with n= 18 per level across the study.

      3.6 Freeze-thaw stability

      The concentration of each calibrator and QCs (low and high) for both sets of freeze-thaw cycles (from -80°C to room temperature and -20°C to room temperature) were measured and plotted for each run. The data obtained was first observed in JMP Pro 15.2 for outliers. For all six freeze-thaw cycles of both freezing temperatures, the concentration per calibrator level was first compared using a two-tailed Student's paired t-test (α= 0.05) for the technical replicates. No significant difference was observed in the measured concentrations for either temperature across all calibrator levels. In addition, we used the connecting letter report generated by JMP to confirm that the measured calibrators were indeed significantly different from each other and the blank matrix. This is especially important in the context of the lowest-concentration calibrators (L1 and L2) that are only slightly above the background noise level. The connecting letter report indicated the significant difference among all calibrator levels and blanks and showed no significant difference for the same level across runs. This concludes that there is no degradation due to cryopreservation and thawing on the analytes and the same batch of manufactured calibrators can be used for multiple assays over time. The average concentration per calibrator level, per freeze-thaw cycle run is available on Supplemental Fig. S3.

      3.7 Sample carry-over

      To assess sample carryover, we first established a baseline signal for L1 (lowest calibrator level), followed by one L6 (highest calibrator) injection, followed by 3 immediate injections of L1. MRM analyses of both analytes showcased no statistically significant deviation in the observed L1 measurements from the L1 baseline, indicating that there was no carryover effect in any of the measurements carried out (data spread shown in Supplemental Fig. S4). Minimal carryover effect is desirable particularly if this method is to be implemented at a clinical setup. Performing blank washes after each patient sample injection would mean that fewer samples can be tested per plate to accommodate the wash wells. Our method integrates autosampler Needle Wash Solvent Chemistry (as described in the methods section), thus mitigating the sample carryover. This method can be easily transferred for many patient samples tests without the need for any blank insertion after each sample injection.

      3.8 Kit stability

      For our 14-day accelerated stability study (488 days extrapolated), we ran two plates per temperature point and duplicates per point. The data obtained was first converted to % change (in relation to Day= 0 samples) and then normalized. For both 20C and +30C stress temperatures, it was difficult to obtain a conclusive trend for either analyte over the 14-day period. There was a significant difference between the concentration obtained for D=0 and D=14; however, there was no conclusive result for when this decline/degradation began to happen. To take a closer look at this, we ran the D=0, D=2, D=4, D=6 samples for all calibrators for 20C stress temperature in triplicate. For this run, we looked at the raw concentration values across the 6-days, as well as the % change and the normalized data. This data was then extrapolated to obtain a degradation equation that could be used to plot a trendline. The data obtained from our initial study was plotted against the predicted trendline from the extrapolated data. It was interesting to observe that while the extrapolated trend predicted a steady 2% decrease in cortisol concentration and a 4% decrease in cortisone concentration, for each day of the accelerated study, the true measured data (in duplicates, across 14-days) exhibited a much slower decline in concentration. Because the predicted values are extrapolated and the experimental values were in duplicates, Fig. 4 shows no error bars. We then performed a Tukey-Kramer HSD analysis to obtain any statistically significant difference for each level across the six days (measured in triplicates), hoping to reach a conclusion for the stability of the designed ‘kit’. The generated connected letter report showed no significant difference in the means of the concentration measured (for both analytes) across the 4-days, stating that the analytes are stable for 8.5 months (256 days) at 80C storage. It is important to note that the reported decline is well-within our 20% tolerance for precision and accuracy, and since it is an accelerated study, the stability likely extends beyond the time points tested.
      Fig. 4
      Fig. 4JANUS® G3 Workstation (left) versus human (right) standard curves for cortisol and cortisone (qualifier and quantifier); n=3 per level per plate. The standard curve obtained from JANUS plating showed less spread in the data points compared to the human plate, although the assay linearity was measured at 0.99 and above for both.

      3.9 Automation efficacy, precision, and accuracy

      Owing to the ease of sample preparation, we were able to quickly adapt the sample preparation protocol to the JANUS® G3 workstation. The established automation protocol demonstrated the capability of completing a precision study on 48 samples in 4.93 minutes, compared to the 6-minute of human plating (human intervention and on-deck actions are illustrated in Supplemental Fig. S5). The reported R-Square value for the linear regression for an average of 3 automated runs was ≥0.99, demonstrating that it possesses the required assay linearity. Moreover, automated sample runs reported a considerable improvement in precision (%CV), showcasing a decreased spread in data points across the standard curve (particularly for the higher-concentration calibrators). Since we have already established the overall linearity and accuracy of this assay, we adhered to the use of data points spread in standard curves as a measure of improved precision (Fig. 5). The automated procedure made the entire assay just above 3-minutes for plating, compared to the human 6-minutes plating. It is important to note that human plating required re-capping the calibrator, QC, and sample vials, to prevent contamination, whereas all the vials could be kept open during the entire duration of plating on the JANUS deck. This action will prove particularly advantageous when translating this method at a high throughput laboratory or clinical setting. Because of the already established assay precision and accuracy, the automation step can be designed for single calibrator injections, leaving the user 81-wells open for at least 81-unknown samples testing in singlicate. Given our study utilizing a small volume (20 µL) sample, small human error in pipetting can magnify, propagate and have a significant impact on the assay precision. That is one avenue where automation greatly helps due to its ability to accurately dispense small volume samples, minimizing error propagation.
      Fig. 5
      Fig. 5Quantifier (top) and Qualifier (bottom) concentrations measured against the theoretical concentrations for cortisol-cortisone. The measured concentrations were used in the generation of the standard curves for all further tests.

      4. Conclusions

      The presented work successfully establishes an optimized LC-MS/MS method for the simultaneous detection of free cortisol-cortisone in saliva. Moreover, a quantification kit for the diagnostic evaluation of cortisol-cortisone was developed. Based on a wide number of assessments performed over the course of this work, the proposed LC-MS/MS method is rapid, sensitive, and precise. The methodology showcased the stability, linearity, and simplicity required for theoretical clinical use. When compared with previous reports, this method and quantification kit additionally showcased a heightened simplicity in the sample preparation process, owing to the lack of expensive and complex phase separation processes required in previous literature. Overall, this allowed for the full automation of the sample preparation process, furthering the usability of this work in the clinical setting. Pending further validation, this quantification kit may be used to provide valuable information in the clinical diagnostic realm for therapeutic monitoring of cortisol cortisone to elucidate potential adrenal dysfunctions in patients suffering from CAH, Cushing's Syndrome, as well as other endocrine disorders.

      Declaration of interests

      The authors declare the following financial interests/personal relationships which may be considered as potential competing interests:
      Anubhav Tripathi reports financial support was provided by PerkinElmer Inc. Anubhav Tripathi reports a relationship with PerkinElmer Inc that includes: consulting or advisory.

      Acknowledgments

      We would like to thank Kathryn Sherlock for helping with the automation troubleshooting, and Rudra Maharaj for his valuable insight during the initial phase of this study. We would also like to thank Celeste Welch and Lindsay Schneider for their valuable scientific feedback throughout the study. A.T. is a paid scientific advisor/consultant and lecturer for PerkinElmer.

      Appendix. Supplementary materials

      References

        • Timmermans S.
        • Souffriau J.
        • Libert C.
        A general introduction to glucocorticoid biology.
        Front Immunol. 2019; 10: 1545
        • Stephens M.A.C.
        • Wand G.
        Stress and the HPA axi: Role of glucocorticoids in alcohol dependence.
        Alcohol Res. 2012;
        • Ehlert U.
        • Gaab J.
        • Heinrichs M.
        Psychoneuroendocrinological contributions to the etiology of depression, posttraumatic stress disorder and stress-related bodily disorders: the role of the hypothalamus–pituitary–adrenal axis.
        Biol Psychol. 2001; 57: 141-152
        • Nicolaides N.C.
        • Galata Z.
        • Kino T.
        • Chrousos G.P.
        • Charmandari E.
        The human glucocorticoid receptor: molecular basis of biologic function.
        Steroids. 2010; 75: 1-12
        • Ramamoorthy S.
        • Cidlowski J.A.
        Corticosteroids: mechanisms of action in health and disease.
        Rheum Dis Clin. 2016; 42: 15-31
        • Kuo T.
        • McQueen A.
        • Chen T.-C.
        • Wang J.-C.
        Regulation of glucose homeostasis by glucocorticoids.
        Glucocorticoid Signal. 2015; : 99-126
        • Exton J.
        Regulation of gluconeogenesis by glucocorticoids.
        Monogr Endocrinol. 1979; 12: 535-546
        • Olgemöller B.
        • Roscher A.A.
        • Liebl B.
        • Fingerhut R.
        Screening for congenital adrenal hyperplasia: adjustment of 17-hydroxyprogesterone cut-off values to both age and birth weight markedly improves the predictive value.
        J Clin Endocrinol Metabol. 2003; 88: 5790-5794https://doi.org/10.1210/jc.2002-021732
        • New M.I.
        Diagnosis and management of congenital adrenal hyperplasia.
        Annu Rev Med. 1998; 49: 311-328
        • Yamamoto T.
        Latent adrenal insufficiency: concept, clues to detection and diagnosis.
        Endocr Pract. 2018; 24: 746-755
        • Mitchell I.C.
        • et al.
        Subclinical Cushing’s syndrome” is not subclinical: improvement after adrenalectomy in 9 patients.
        Surgery. 2007; 142 (900-905. e901)
        • Do Yup Lee E.K.
        • Choi M.H.
        Technical and clinical aspects of cortisol as a biochemical marker of chronic stress.
        BMB reports. 2015; 48: 209
        • Seger C.
        • Salzmann L.
        After another decade: LC–MS/MS became routine in clinical diagnostics.
        Clin Biochem. 2020; 82: 2-11
        • Jung C.
        • Greco S.
        • Nguyen H.H.
        • Ho J.T.
        • Lewis J.G.
        • Torpy D.J.
        • Inder W.J.
        Plasma, salivary and urinary cortisol levels following physiological and stress doses of hydrocortisone in normal volunteers.
        BMC Endocrine Disord. 2014; 14: 1-10
        • Tudela E.
        • Muñoz G.
        • Muñoz-Guerra J.A.
        Matrix effect marker for multianalyte analysis by LC-MS/MS in biological samples.
        J Chromatogr B. 2012; 901: 98-106
        • Papanicolaou Dimitris A.
        • et al.
        Nighttime salivary cortisol: a useful test for the diagnosis of Cushing's syndrome.
        J Clin Endocrinol Metabol. 2002; 87: 4515-4521
        • Bacila Irina
        • et al.
        Measurement of salivary adrenal-specific androgens as biomarkers of therapy control in 21-hydroxylase deficiency.
        J Clin Endocrinol Metabol. 2019; 104.12: 6417-6429
        • Pasquali Renato
        • et al.
        Cortisol and ACTH response to oral dexamethasone in obesity and effects of sex, body fat distribution and dexamethasone concentrations: a dose-response study.
        J Clin Endocrinol Metabol. 2002; 87.1: 166-175
        • Ospina Naykky Singh
        • et al.
        ACTH stimulation tests for the diagnosis of adrenal insufficiency: systematic review and meta-analysis.
        J Clin Endocrinol Metabol. 2016; 101.2: 427-434
        • Hannibal K.E.
        • Bishop M.D.
        Chronic stress, cortisol dysfunction and pain: a psychoneuroendocrine rationale for stress management in pain rehabilitation.
        Phys Ther. 2014; 94: 1816-1825
        • Bakusic J.
        • De Nys S.
        • Creta M.
        • Godderis L.
        • Duca R.C.
        Study of temporal variability of salivary cortisol and cortisone by LC-MS/MS using a new atmospheric pressure ionization source.
        Sci Rep. 2019; 9: 1-12
        • Magda B.
        • Dobi Z.
        • Mészáros K.
        • Szabó É.
        • Márta Z.
        • Imre T.
        • et al.
        Charged derivatization and on-line solid phase extraction to measure extremely low cortisol and cortisone levels in human saliva with liquid chromatography–tandem mass spectrometry.
        J Pharm Biomed Anal. 2017; 140: 223-231
        • Antonelli G.
        • Ceccato F.
        • Artusi C.
        • Marinova M.
        • Plebani M.
        Salivary cortisol and cortisone by LC–MS/MS: validation, reference intervals and diagnostic accuracy in Cushing’s syndrome.
        Clin Chim Acta. 2015; 451: 247-251
        • Fustinoni S.
        • Polledri E.
        • Mercadante R.
        High-throughput determination of cortisol, cortisone and melatonin in oral fluid by on-line turbulent flow liquid chromatography interfaced with liquid chromatography/tandem mass spectrometry.
        Rapid Commun Mass Spectrom. 2013; 27: 1450-1460
      1. Niewoehner, C. B. (2004). 5. In Endocrine pathophysiology (Second, pp. 85–105). essay, Hayes Barton Press.

        • Stewart P.M.
        • Corrie J.E.T.
        • Shackleton C.H.L.
        • Edwards C.R.W.
        Syndrome of apparent mineralocorticoid excess. A defect in the cortisol-cortisone shuttle.
        J Clin Invest. 1988; 82: 340-349
        • Casals G.
        • Hanzu F.A.
        Cortisol measurements in Cushing’s syndrome: immunoassay or mass spectrometry?.
        Ann Labor Med. 2020; 40: 285-296
        • Chabre O.
        The difficulties of pseudo-Cushing’s syndrome (or “non-neoplastic hypercortisolism”).
        Ann Endocrinol. 2018; 79: 138-145
        • Raff H.
        • Phillips J.M.
        Bedtime salivary cortisol and cortisone by LC-MS/MS in healthy adult subjects: evaluation of sampling time.
        J Endocr Soc. 2019; 3: 1631-1640
        • Erickson D.
        • Singh R.J.
        • Sathananthan A.
        • Vella A.
        • Bryant S.C.
        Late-night salivary cortisol for diagnosis of Cushing’s syndrome by liquid chromatography/tandem mass spectrometry assay.
        Clin Endocrinol. 2012; 76: 467-472